Consumables and Spare Parts
There are only two main consumables for the APC-2 Particle counter, sample cups and “O” ring seals for the in-line filter.
...
-
Instrument Errors – Drive Board on 402 Unit
Issue: 402 Unit does not work. You may experience an intermittent green light, and/or intermittent fan movement.
Generally this is a failed power transistor on the Drive board. Ty...
-
Instrument Errors - Baumer Level Sensor Incorrect Readings
Issue: Odd height check readings from the Baumer sensor using the height disk or incorrect solvent dilutions.
Check to see if there is any oil buildup at the end of the sensor snout. &nb...
-
Instrument Errors - Particle Counter Reports Zero Counts
When the particle counter reports zero counts, it normally means that the LDS sensor is dirty but it will also give zero counts if the sampling syringe does not fill completely and a measurement take place with air and not sample flowing through t...
-
Sensor Calibration - Materials & Procedure
The CINRG CS-APC-2 can be calibrated on-site by a laboratory technician with the correct hardware and standards in place by following the calibration procedure outlined in t...